This week, the team had several tasks on their mind: troubleshoot
the circuit (which last week had been soldered in the device), fix broken
spindles, get the entire system working, and prepare for the midterm design
review.
Early in the week,
Wyatt and Adam localized the issue of the circuit malfunction to
misunderstanding the pins of a mechanical switch used three times in the
device. After re-soldering the connections to the pins, the circuit read
the correct output signals, measured on an oscilloscope. A formal circuit diagram was drawn up. Now the only thing left
to do with the electronics is to tap leads from the electronics box to connect
to the EMG electrodes on the mother’s belly. These will be the one disposable part
of the design.
| Figure1. Soldered electronics box. |
| Figure2. Formal circuit diagram of T.A.C.O. electronics drawn in Multisim. |
| Figure3. Motor with new spindles functioning off the power supply without tension. |
Elated, we connected the entire system, and turned on the device.
With care, we began to turn the manual potentiometer for each motor and were
dismayed at the sound of snapping PLA! The timing belt was also still slipping
periodically. Some analysis shows this could have been predicted. The issues
were now isolated to the following: the set screws were too long, and stuck
out, causing the walls of the housing to obstruct their path as they spun with
the timing belt pulley; the timing belt pulley was still not concentric; and
the size of the PLA spindle end was unable to handle the stresses developed
when the motor was running with resistance. The system has been redesigned to
include a single metal axle for the spindle, which will extend from the band,
through the timing belt, into a bearing that will be connected to a mechanical
support for the timing belt system. The redesign will be drawn up in SolidWorks
next week.
In addition to all this, the converter had a failed print. This
showed us the cross section of the part, revealing unexpected geometry. Joel
worked with CBU’s resident 3D printing expert to redesign the converter to
account for this artifact of the previous design. Ruthie contacted Sierra-Medical about obtaining a cow uterus for testing, and logistics and detailed plans for this process are still being worked out. Wyatt redesigned the electronics box to better organize the electronics.
Preparation for the midterm design presentation was primarily remote,
because the team did not have coinciding schedule’s this week. The presentation
is taking place this afternoon, Monday, February 27th, at Bourns Laboratory at 2:45 p.m.